Super-Resolution Methods in MRI: Can They Improve the Trade-Off Between Resolution, Signal-to-Noise Ratio, and Acquisition Time?
E. Plenge, D. H. J. Poot, M. Bernsen, G. Kotek, G. Houston, P. Wielopolski, L. van der Weerd, W. J. Niessen, E. Meijering
Magnetic Resonance in Medicine, vol. 68, no. 6, December 2012, pp. 1983–1993
Improving the resolution in magnetic resonance imaging (MRI) comes at the cost of either lower signal-to-noise ratio (SNR), longer acquisition time, or both. This study investigates whether so-called super-resolution reconstruction (SRR) methods can increase the resolution in the slice selection direction and, as such, are a viable alternative to direct high-resolution acquisition in terms of the SNR and acquisition time trade-offs. The performance of six SRR methods and direct high-resolution acquisitions was compared with respect to these trade-offs. The methods are based on iterative back-projection, algebraic reconstruction, and regularized least squares. The algorithms were applied to low-resolution data sets within which the images were rotated relative to each other. Quantitative experiments involved a computational phantom and a physical phantom containing structures of known dimensions. To visually validate the quantitative evaluations, qualitative experiments were performed, in which images of three different subjects (a phantom, an ex-vivo rat knee, and a post-mortem mouse) were acquired with different MRI scanners. The results show that SRR can indeed improve the resolution, SNR, and acquisition time trade-offs compared with direct high-resolution acquisition.
Copyright © 2012 the authors. Published version © 2012 John Wiley & Sons. Personal use of this material is permitted. However, permission to reprint or republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works, must be obtained from the copyright holder.