3D Inversion Recovery MEMRI Using Super-Resolution Reconstruction
E. Plenge, D. S. Poole, D. H. J. Poot, W. J. Niessen, L. van der Weerd, E. Meijering
Hot Topics in Molecular Imaging (TOPIM 2012)
Held in Les Houches, France, April 15–20, 2012
Introduction: Manganese-enhanced MRI (MEMRI) is increasingly used in small animal imaging studies for the unique T1-contrast it provides. Often, due to time limitations, but not only, T1-weighted gradient-echo (GE) sequences are used to monitor the manganese uptake, but spin-echo (SE) sequences offer the advantage of being less susceptible to field inhomogeneities. Also, a better contrast between the regions with high and low manganese uptake is obtainable with inversion recovery spin-echo (IR-SE) sequences, but high-resolution 3D imaging in this mode is unfeasible in-vivo due to extensive acquisition times. Such images would, however, be of great value for visualization and analysis purposes. Here we investigate whether super-resolution reconstruction (SRR) methods may provide for this. Such methods allow to create a high-resolution 3D image from a set of low-resolution 2D slice-stacks that each sample the field-of-view in a distinct fashion. For sequences such as IR-SE, which cannot practically be acquired in 3D, SRR can be a useful alternative.
Methods: We constructed a 2% agar phantom with compartments containing manganese concentrations characterized by T1 values of 650, 720, 900, and 1000 ms. A set of 12 IR-SE slice-stacks (TE/TI/TR = 8/650/2700 ms, 1 excitation, 30 slices of 1 mm thickness, in-plane isotropic resolution of 0.2 mm, acquisition time per stack = 6 min) of the phantom was acquired, rotating the stacks in increments of 180/12 degrees around the frequency encoding axis. From these, an image was reconstructed on an isotropic 0.2 mm resolution grid using SRR. For comparison, a 3D GE T1W image (RARE 3D T1W-GE: TE/TR = 3.2/30 ms, flip angle = 30, 4 excitations, isotropic resolution = 0.2 mm, acquisition time = 78 min) was acquired. All images were obtained with a 7T Bruker Pharmascan horizontal bore system. SRR was performed on the IR-SE data-set using a Tikhonov-regularized reconstruction scheme minimizing aliasing and spectral distortions.
Results: Contrast and contrast-to-noise ratio (CNR) between two compartments with manganese concentrations characterized by T1-values of 650 and 900 ms, respectively, was measured in the SRR IR-SE and GE images. In the GE image the contrast was 1.43, the CNR was 20.13, and the acquisition time was 78 min. In the SRR IR-SE image the contrast was 1.87, the CNR was 21.97, and the total acquisition time for 12 slice-stacks was 72 min, a time compatible with in-vivo mouse experiments. A 3D IR-SE acquisition with equivalent contrast and resolution settings was estimated to take ~18 hours. On visual inspection the resolution of the SRR IR-SE image was comparable to that of the GE image.
Conclusions: We have shown that SRR can be applied to obtain high-resolution isotropic 3D MEMRI images from 2D slice stacks, in a time-frame compatible with in-vivo experiments, and that such images offer better contrast than the widely used GE images.
Copyright © 2012 the authors. Published version © 2012 European Society for Molecular Imaging. Personal use of this material is permitted. However, permission to reprint or republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works, must be obtained from the copyright holder.