Simulation-Based Validation of an OCT-Microperimetry Structure-Function Analysis Pipeline in Epiretinal Membrane

A. Jessup, M. M. Rahaman, T. Hong, L. Phan, E. Meijering, A. Sowmya, A. Chang

Annual Meeting of the Association for Research in Vision and Ophthalmology (ARVO 2026)
Held in Denver, CO, USA, May 3–7, 2026


Purpose: Epiretinal membranes (ERM) produce structural macular tissue changes but their relationship to altered vision and retinal function is not fully understood. This study developed and validated an artificial intelligence (AI) algorithm to analyze the relationship between structure and function by correlating paired synthetic optical coherence tomography (OCT) structural maps and microperimetry pointwise macular sensitivity measurements.

Methods: We retrospectively identified eyes with ERM and paired retinal sensitivity measures (CenterVue MAIA). Real microperimetry data (37 macular test locations, sensitivity in dB) were extracted from MAIA results. Synthetic OCT structural maps were generated on a 512 × 512 grid using an analytic function of radial distance from the image center (300 − 20R) and added sinusoidal spatial variation (30*sin(X)*cos(Y)). Maps were perturbed with Gaussian noise, clipped to 100–400 μm, and smoothed (Gaussian filter, σ = 2). A 37-point MAIA grid was registered to the simulated OCT frame, and local structural values were sampled at each test location. Pointwise Pearson and Spearman correlations between simulated structure and real sensitivity were computed per eye.

Results: Nine eyes with ERM and usable MAIA reports were processed successfully, yielding 333 pointwise thickness-sensitivity pairs (37 macular test locations per eye). Across the cohort, per-eye pointwise Pearson correlations between simulated OCT-like thickness and MAIA sensitivity showed predominantly weak-to-moderate associations, with coefficients ranging from r = 0.006 to r = 0.542 (mean r = 0.254 ± 0.180; median r = 0.213). Two of nine eyes (22%) demonstrated statistically significant correlations with p < 0.05, including one eye (11%) with p < 0.01. The strongest observed correlation reached r = 0.542. The remaining eyes showed weak relationships. The automated summary visualization showed the distribution and variability of structure-function coupling across eyes.

Conclusions: The study showed an automated AI analysis and visualization framework for pointwise OCT-microperimetry structure-function analysis in ERM using simulated OCT structural maps and real MAIA sensitivity data. The model may be used to understand biological structure-function relationships. Future research will apply the model to true measured retinal thickness maps, enabling coupled quantitative assessment of structure-function.


Copyright © 2026 the authors. Published version © 2026 Association for Research in Vision and Ophthalmology. Personal use of this material is permitted. However, permission to reprint or republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works, must be obtained from the copyright holder.