Super-Resolution Reconstruction in Molecular MRI — A Feasibility Study

E. Plenge, D. H. J. Poot, M. Bernsen, G. Kotek, G. Houston, P. Wielopolski, M. Halberstadt, W. J. Niessen, E. Meijering

European Molecular Imaging Meeting (EMIM 2011)
Held in Leiden, Netherlands, June 19–21, 2011

Update

Introduction: With its excellent contrast capabilities MRI is a prominent modality in the field of molecular imaging. In practice, however, the spatial and temporal resolutions of an MR image are severely limited by noise and timing considerations. Therefore, quantification and characterization of events at the molecular level, using MRI, will generally require careful balancing of the trade-off space spanned by resolution, signal-to-noise ratio (SNR) and image acquisition time. High resolution, high SNR and low acquisition time are desirable properties of an experiment, and for a given imaging protocol improvement of one parameter is only achievable at the expense of one or two of the others.

Super-resolution reconstruction (SRR) is the process of producing a high-resolution (HR) image from a sequence of low-resolution (LR) images, where each LR image transforms and samples the HR scene in a distinct fashion. In this study, we investigate whether SRR, employed as a post-processing step, can advantageously modify the MRI trade-off space.

Methods: Experiments were performed using an FSE sequence on a 3T GE clinical scanner. A data set was acquired, containing 24 LR images of anisotropic resolution of 1 × 1 × 4 mm of a water-filled plastic phantom. The field of views (FOV) of these images were rotated around the phase encoding axis in increments of 180/24 degrees. Isotropic HR images were reconstructed from various numbers of LR images on a 1 × 1 × 1 mm grid using SRR. Effective resolution and SNR were measured on the reconstructed images and plotted against the number of LR images used for the reconstruction, in effect denoting the acquisition time. These plots reveal the relations between resolution, SNR and acquisition time.

To compare reconstructed HR images with direct HR acquisitions, a set of six HR images of 1 × 1 × 1 mm resolution, all with identical FOV, was acquired. Averaging one to six of these, the resolution and SNR of direct HR acquisitions could be measured versus the number of averages, in effect the imaging time, enabling comparison with the reconstructed images. Furthermore, direct acquisitions of isotropic resolutions (1.2–1.8 mm) were made. The SNR and effective resolution measured in these were extrapolated in the time dimension according to well-known theoretical relationships.

Results: Contrary to previous results by others, our experiments show that SRR cannot surpass the isotropic resolution of direct acquisition. However, we find that the SRR images achieve a resolution that is only slightly worse than that of the direct acquisition, while having significantly higher SNR.

Conclusions: Using SRR, the trade-off between resolution, SNR and acquisition time can be manipulated, and better trade-offs can be achieved than by direct acquisition. The potential of SRR to improve actual quantitative analyses in in-vivo molecular imaging experiments is yet to be explored.


Copyright © 2011 the authors. Published version © 2011 European Society for Molecular Imaging. Personal use of this material is permitted. However, permission to reprint or republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works, must be obtained from the copyright holder.